EEExxxppprrreeessssss 3 Primary Tube Centrifuge Operator’s Manual A Division of IRIS International, Inc.
6 2. Slide the rotor onto the shaft and push down gently to seat the drive pins in the holes. 3. Place the retaining nut on the shaft and tighten s
7 Unloading 1. Upon completion of the cycle, the rotor decelerates to a complete stop in 30 seconds or less, a signal beep will sound, and the
8 Section 4 Maintenance Overview Iris Sample Processing recommends that instrument operators perform periodic inspections and preventative maintenan
9 Troubleshooting No power light ↓ Is the line cord completely plugged into the wall? → No→ Plug line cord into the wall. ↓ Yes ↓ Is the w
10 Service Refer all service to Iris Sample Processing or an Authorized Repair Facility. 1-800-782-8774 Instrument fuse(s) are located on the rea
11 Section 5 Appendix Specifications Product Number SSX3 Model Number M502-22 Spin Parameters 2 min: 120 seconds @ 7200 rpm / 4440 xg 3 min: 180
12 References 1. CLSI. “Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline-Second Edition.” CLSI docume
13 Iris Sample Processing Warranty Iris Sample Processing, a Division of IRIS International, Inc. warrants that the instruments shall be free from de
Operator’s Manual Express 3 Primary Tube Centrifuge Model Number M502-22 For in-vitro diagnostic use Product Number SSX3 - Express 3 for 100-24
iTable of Contents How to use this manual…………………………………………..……….………… ii Section 1 ……………………………………………………………..……….……….. 1 Unpacking and Installation .
iiHow to use this manual This manual along with information contained on product labels and in package inserts should provide you with all the info
1 Section 1 Unpacking and Installation Inspect Packaging The Express 3 and its accessories are delivered in a single carton. If the centrifuge or a
2 Section 2 System Overview Principle and Intended Use For in-vitro diagnostic use to produce the rapid separation of whole blood contained in origin
3 Operator Controls Start The Start button initiates the selected cycle. Stop/Open The Stop/Open button interrupts the cycle and releases the c
4 Accessories Product No. Description SV02 Inserts for 3 mL tubes (10.25 x 64 mm) 4/pk. SV03 Inserts for 2 mL tubes (10.25 x 47 mm or BD Micro
5 Section 3 Operating Instructions Opening and Closing the Cover An electronic operated cover interlock mechanism prevents operation if the cover