Iris-sample-processing Statspin® MP Multipurpose Centrifuge Manuel d'utilisateur Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 19
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
13
Appendix
Appendix A - Specifications
Product No. SSMP (supplied with 2 rotors, RT12 and RH12)
SSMP-1 (supplied with 1 rotor, RT12)
SSMP-2 (supplied with sample of PlasmaRotors)
Model No. M901
Cycles/Speeds
Urine 9,800 (3,900 x g); 45 seconds
Hematocrit 16,000 (13,700 x g); 120 seconds
Tube Rotor 15,800 (12,000 x g); 30 seconds
PlasmaRotor
®
20,000 (11,200 x g); 30 seconds
LipoClear
®
15,800 (12,000 x g); 95 seconds
Coagplasma 20,000 (11,200 x g); 120 seconds
Acceleration Time Approximately 6 seconds
Deceleration Time Approximately 10 seconds
Electrical 24 Volts, DC, 1.7 amp. Includes switching power supply for 100-240
VAC, 50/60 Hz
Dimensions Diameter 6.6”/ 16.25 cm
Height 6.3”/13.2 cm
Weight 5.5 lbs/2.5 kg
Environmental Indoor use
Altitude up to 2000m
Temperature 5ºC to 40ºC
Maximum relative humidity 80% for temperatures up to 31ºC decreasing
linearly to 50% relative humidity at 40ºC
Main supply voltage fluctuations not to exceed +/- 10% of the nominal
voltage
Transient over-voltages according to installation category II
Pollution degree 2